GE Healthcare – Lunar DPX IQFull Body Bone Densitometer, The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better. service manual for lunar dpx. Sat, 20 Oct GMT service manual for lunar dpx pdf – DPX Series. Service repair lunar dpx iq service manual – free. Frank’s Hospital Workshop. – service manual lunar dpx nt LUNAR DPX Operators . Manual, lunar dpx iq service manual pdf Gluer.

Author: Dougar Arashigami
Country: Cambodia
Language: English (Spanish)
Genre: Health and Food
Published (Last): 24 December 2013
Pages: 494
PDF File Size: 2.71 Mb
ePub File Size: 13.87 Mb
ISBN: 472-5-78850-901-7
Downloads: 55934
Price: Free* [*Free Regsitration Required]
Uploader: Tygorn

You probably have a bad detector if the home positioning is good. Remove Subscription Confirm Remove. We have both new and used options ljnar fit your budget. Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment.

Sign In or Register. This site uses cookies. My Bench Order History Sign out. The language is lumar problem, actually is perfect. If you would like a quote feel free to email me at ben dexasolutions.

service manual for the lunar DPX-IQ

We too stock serviice detectors. Just remember that the detector is not properly responsive during QC and exams as well. This site uses cookies.

  BATELCO ANNUAL REPORT 2011 PDF

Please review our Privacy Policy for more details. Advertise Increase Your Brand Awareness. Also, as was mentioned, this is a part you do not need any specialty tools to replace if the alignment has not changed. This is a very common failure on any of the DPX models. Forums Documents Parts Videos News.

Tissue value 1, Waiting for a prompt answer. Hi Ben, I would like get you help. Having worked for the Lunar business for many years we have seen these failures numerous times. Add to My Bench. I have a problem with this team is that everything works, from the source to the Oink card, but does not open the solenoid of the collimator.

The DPX IQ provides the speed of entry level fan beam bone densitometer systems with jq image quality.

Service manual for GE Lunar DPX-NT densitometer

You can sometimes get past this by using the x-ray dx mode to saturate the detector. By continuing to browse the site you are agreeing to our use of cookies. Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment. The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better image quality. If you like i find you a reloaded detector.

Contact Us Now – Log in or Register to rate and post a comment. On the momento the bmd is in 1, L2-L4,normal ranging 1, to 1,the last reading before of the fail was 1, Hi Adriano, As mentioned, if nothing has changed recently with alignment, then your detector has deteriorated.

  EPSON TW6100 MANUAL PDF

Tengo un problema con este equipo, resulta que todo funciona, sdrvice la fuente, hasta la tarjeta Oink, pero no abre el solenoid del colimador. mannual

GE Healthcare – Lunar DPX IQ Community, Manuals and Specifications | MedWrench

This is what happens when you run the QA multiple times back to back. Remove Subscription Confirm Remove. Add to My Bench. You can adjust your Community Subscriptions in Settings. If active manually if the solenoid is activated, similarly, if I activate the switch, achievement cheat the system and I do not make the mistake, but if only the active from the system, it does nothing.

Greetings and thanks for your help Reply. You can adjust your Community Subscriptions in Settings.

You likely have a bad detector. You can add Community Subscriptions in the search bar that says “Subscribe to more communities